Number of wells | 96 |
Principle | Direct Sandwich |
Assay format | Semi-Quantitative |
Total incubation time | 105 minutes |
Sample type | Serum or Plasma |
Sample volume | 10 μL |
Reading wavelength | 450nm |
Storage | 2-8°C |
Certification | Taiwan EUA 1100000263 / CE |
5 serum samples collected from RT-PCR confirmed COVID-19 patients were tested. 139 normal healthy patients with samples collected before COVID-19 outbreak were tested. The results are as follows:
PERFORMANCE | Test Positive | Test Negative |
Confirmed Positive | 5 | 0 |
Confirmed Negative | 0 | 139 |
Inter-assay precision CV | < 10% | Inter-assay precision CV | < 20% |
The diagnostic sensitivity is 100%;The diagnostic specificity is 100%
Negative (n=89) | Positive (n=5) | ||||||
Company A | Company A | ||||||
Pos | Neg | total | Pos | Neg | total | ||
GBC | Pos | 0 | 0 | 0 | 4 | 1 | 5 |
Neg | 2 | 87 | 89 | 0 | 0 | 0 | |
total | 2 | 87 | 89 | 4 | 1 | 5 | |
Sensitivity | GBC | 100% | |||||
Company A | 80% | ||||||
Specificity | GBC | 100% | |||||
Company A | 97.7% |
Catalogue No. | Product |
4ECO001 | GB SARS-COV-2 Ab ELISA - 96T |
4ECO002 | GB SARS-COV-2 Ab ELISA - 480T |